Atherosclerosis and Lipoproteins |
From The Methodist Hospital Research Institute & The Methodist DeBakey Heart Center (J.F.G., E.J.L., D.G.S., A.E.R., G.L.K.), Houston, Tex; Department of Mathematics (D.W.B.), University of Houston, Houston, Tex; Baylor College of Medicine (H.K.W.), Houston, Tex; The Corbic Research Foundation-HMUA (C.L.A.), Envigado, Colombia; Instituto de Ciencias de la Salud (A.B.), Medellin, Colombia; Department of Pathology (R.B.), The Methodist Hospital, Houston, Tex.
Correspondence to Juan F. Granada, MD, The Methodist Hospital Research Institute, The Methodist DeBakey Heart Center, Houston, TX 77584. E-mail jgranada{at}tmh.tmc.edu
| Abstract |
|---|
|
|
|---|
Methods and Results Coronary lesions were induced by injecting liposomes containing human oxidized low-density lipoprotein into the adventitia of the arteries. IVUS-VH imaging was performed in vivo at 8.2±1.6 weeks after injection. A total of 60 vascular lesions were analyzed and compared with their correspondent IVUS-VH images. Correlation analysis was performed using linear regression models. Compared with histology, IVUS-VH correctly identified the presence of fibrous, fibro-fatty, and necrotic tissue in 58.33%, 38.33%, and 38.33% of lesions, respectively. The sensitivity of IVUS-VH for the detection of fibrous, fibro-fatty, and necrotic core tissue was 76.1%, 46%, and 41.1% respectively. A linear regression analysis performed for each individual plaque component did not show strong correlation that would allow significant prediction of individual values.
Conclusions In a porcine model of complex coronary lesions, IVUS-VH was not accurate in detecting the relative amount of specific plaque components within each individual corresponding histological specimen.
Coronary lesions were developed by adventitial injection of lipid-rich liposomes in porcine coronary arteries. IVUS-VH analysis was performed at 8.2±1.6 weeks after injection. Linear regression models showed that IVUS-VH was not accurate in detecting the relative amount of specific plaque components within each individual corresponding histological specimen.
Key Words: animal model intravascular ultrasound vulnerable plaque
| Introduction |
|---|
|
|
|---|
Virtual histology is an intravenous ultrasound (IVUS)-based technology that studies the spectral analysis of the radio frequency signals backscattered from the plaque during IVUS imaging and reconstructs the morphology of the plaque through color-coded maps.11 Although this technique has been validated in vitro and ex vivo in human coronary arteries1114 and preliminary clinical experience has been published,15,16 little data have been published on histological correlates of images acquired in vivo.17 In the present study, we describe the histological characteristics of complex coronary lesions analyzed in vivo using intravascular ultrasoundvirtual histology (IVUS-VH) in a porcine model of disease.
| Methods |
|---|
|
|
|---|
IVUS Imaging Analysis
All IVUS images were acquired using a 20-MHz Volcano Eagle EyeTM IVUS catheter (Volcano Therapeutics Inc, Rancho Cordova, Calif). At follow-up, systemic anticoagulation using heparin and intracoronary nitroglycerine was used before imaging. Once the coronary lesion was identified, the IVUS catheter was inserted distal to the lesion and manually pulled back to assess the severity and length of the lesion. The IVUS catheter was then placed distal to a side branch (distal fiduciary landmark site) and automatic pullback was performed at a rate of 1 mm/sec. The location of the IVUS catheter was determined using continuous fluoroscopy during the entire pullback time and by recording anatomic landmarks seen during IVUS imaging. To achieve adequate images, an average of 2 pullbacks per artery was performed and the best play loop was chosen based on imaging resolution and quality. Continuous EKG monitoring was performed during the procedure to gate IVUS frames for analysis. A complete morphometric analysis was performed in 532 independent IVUS frames using standard definitions.19 All measurements were automatically derived from the Volcano Invision Gold imaging system software.
IVUS-VH data were recorded to the imaging system hard drive and then extracted and archived for analysis. The analysis was based on border contour calculation from gray scale. The same tissue maps provided by the software (dark green=fibrous, light green=fibro-fatty, red=necrotic core and white=dense calcium) were used to analyze each independent frame. Once the total length of the lesion was determined, a 20-mm vascular segment containing the vascular lesion was selected for analysis. This segment was then divided in equal 2.0-mm subsections, generating a total of 10 series of cross-sections per vascular segment.
Gray level median of echogenicity analysis was performed by visually and mathematically matching histological slides with the same gray scale IVUS frames used for VH analysis. The gray scale intensities of the plaque (external elastic lamina area lumen area) were recorded and a gray scale histogram produced. The histogram was divided into tertiles corresponding to the lower, mid, and upper echogenic regions within the IVUS image. These regions were then color coded (dark green=fibrous, light green=fibro-fatty, red=necrotic core) and compared with all 3 different tissue types. To determine the sensitivity of detection of the technique, each detected component was compared with the corresponding histological slide after the same methodology used for the IVUS-VH analysis.
All IVUS frames were analyzed independently by 2 observers and were used for comparison with corresponding histological sections. These observers performed IVUS quantitative analyses for each cross section independently (interobserver variability). In case of disagreement, the observers reanalyzed the IVUS image and reached a consensus diagnosis. The observers then repeated a blinded analysis after an interval of 2 weeks (intraobserver variability).
Histological Analysis
At last follow-up, each pig was euthanized, heart was retrieved, and the coronary circulation was anterogradely perfused with 1 L of heparinized normal saline at 100 mm Hg to remove residual blood. Individual coronary arteries were excised and the matching fiduciary distal landmark macroscopically identified. Sequential 2-mm sections were obtained starting at the same anatomic position in which IVUS pullback was started (distal to proximal) in a 20-mm arterial segment. A total of 10 corresponding histological blocks (2-mm-thick) were generated and placed in individual histology cassettes labeling their orientation and landmarks for future reference. The tissue was then preserved frozen on liquid nitrogen (LN2) and sent for histological staining. Serial 5-µm sections were cut from each histological block and stained with hematoxylin and eosin, Oil-red-O, Masson trichrome/elastin stain, and Movat pentachrome. Immunohistochemical staining was performed using a myeloperoxidase-labeled streptavidinbiotin method from Neomarkers (Fremont, Calif) to detect the presence of macrophages. Sections were also immunolabeled with an anti-factor VIII antibody followed by a standard avidin-biotin peroxidase complex assay Vector ABC Elite Kit (Vector Laboratories) to stain for neovessels. Slides were developed with DAB (Zymed Laboratories) and counter-stained with 10% hematoxylin. After staining, 24-bit, full-color digital images were acquired at a magnification to include the entire vessel cross-section (2x) using an Olympus U-CMAD3 camera (Melville, NY) attached to an Olympus BX41 microscope (Melville, NY) at a resolution of 1024x768 pixels. Morphometric analysis was performed using Image-Pro plus 4.5 (Media Cybernetics, Silver Spring, Md). The characterization of each individual morphological component was performed using standard definitions previously published.11,13,20,21 Areas of densely packed collagen were labeled as fibrous and those with significant lipid interspersed in collagen were labeled as fibro-lipidic. Necrotic regions were defined as areas of the plaque not occupied by collagen and containing cholesterol clefts, foam cells, and micro-calcifications. Calcium deposits without adjacent necrosis were identified as calcium. The individual total plaque area and individual plaque components were recorded for each individual histological slide and the sum of all of the components matched to the total plaque area.
Statistical Analysis
The data are presented as mean±SD for parametric data. Standard statistical methods were used by using SigmaStat 3.1 software for comparison between the groups. Correlation between histology and virtual histology values were evaluated by using linear regression models. The degree of agreement between the inter-observer and intra-observer variability were quantified using linear regression models and Bland-Altman plots.
| Results |
|---|
|
|
|---|
In Vivo IVUS-VH Analysis
IVUS was performed in vivo in all the lesions at baseline, 2 weeks, and at euthanization. Both intra-observer and inter-observer variability were minimal and within acceptable limits for the detection of luminal and media adventitial boundaries as shown in Figure 1. At last follow-up, the mean IVUS pullback length analyzed was 27.5±6.2 mm. IVUS classified these lesions as eccentric, echolucent, and with no evidence of calcification. These lesions were consistently non-occlusive (41.3%±8.0% plaque burden) and located in remodeled vessels (remodeling index, 1.01±0.08). The mean vessel area was 16.7±3.0 mm2 and the mean plaque area was 6.9±1.9 mm2. IVUS-VH analysis categorized these lesions as having large amount of fibrous tissue (mean: 55.5%±22.4%) and lesser amounts of fibro-lipidic (mean, 11.4%±10.7%), necrotic (mean, 17.7%±13%), and calcific (mean, 14.4%±16%) tissue.
|
Histological Evaluation of Vascular Lesions
A total of 60 coronary slides were included in the histological analysis. The mean plaque area (0.99±0.59 mm2 versus 0.05±0.03 mm2; P<0.01) and the mean percentage of plaque area (58.9% ± 14% versus 7.5% ± 4.9%; P<0.01) were significantly different compared with noninjected segments. All lesions were classified as eccentric and fibro-lipidic vascular lesions (mean percentage area for fibrous tissue 40.3%±12.8% and 16.8%±8.19% for fibro-lipidic). The presence of lipid deposits colocalized with the presence of tissue macrophages (mean macrophage count per histological slide was 32.7±13.4) and adventitial neovascularization (mean count per plaque 33.1±12.8). There was no evidence of intra-plaque hemorrhage or calcium deposits in any of the lesions (Figure 2).
|
Characterization of Plaque Components by VH
A histologyIVUS-VH correlation analysis was performed in 60 vascular lesions. Because none of the histological sections showed evidence of significant calcification, the correlation analysis was performed for all 3 remaining tissue types. IVUS-VH correctly identified the presence of fibrous tissue in 35 of 60 lesions (58.33%), fibro-fatty tissue in 23 of 60 lesions (38.33%), and necrotic core tissue in 23 of 60 lesions (38.33%). The sensitivity of IVUS-VH for the detection of fibrous, fibro-fatty, and necrotic core tissue was 76.1%, 46%, and 41.1%, respectively. A linear regression analysis performed for each individual plaque component did not show strong correlation that would allow significant prediction of individual values (Figure 3). Similar analysis performed using gray level median of echogenicity demonstrated an improved sensitivity for detection of different plaque components to 86.9%, 79.5%, and 89.7% for fibrous, fibro-fatty, and necrotic core tissue, respectively (Figure 4).
|
|
| Discussion |
|---|
|
|
|---|
Our study demonstrated that conventional gray scale IVUS was capable of identifying and quantifying all vascular lesions at 10 weeks. Using the same IVUS frames used for the IVUS-VH analysis, gray level median of echogenicity analysis was able to improve the characterization of all plaque components with an acceptable degree of accuracy. IVUS-VH analysis of the same images acquired in vivo consistently detected fibrous tissue as the most prevalent tissue type found in these lesions (mean, 55.5%±22.4%). Also, similar proportions of fibro-lipidic (mean, 11.4%±10.7%), necrotic (mean, 17.7%±13%), and calcific (mean, 14.4%±16%) tissue were also detected by IVUS-VH. Although the sensitivity of IVUS-VH for the detection of fibrous tissue was high, the specificity was poor because of high false-positive rates. Specificity of IVUS-VH for detection of fibro-fatty plaque was 80% and sensitivity was 46%. Sensitivity and specificity for detection of necrotic core were both modest at 41.1% and 50%, respectively. In addition, the presence of calcium was consistently detected despite the fact this component was absent in all histological specimens (Figure 5). A linear regression analysis between IVUS-VH and histology showed that in our model, IVUS-VH was not accurate in detecting the area of each specific component within the corresponding histological specimen. Therefore, in our study, sensitivity and specificity of IVUS-VH in detecting different tissue types were not as robust as had been validated in previous ex vivo studies.
|
Our findings may have significant implications in regard to the methodology currently used to validate catheter-based plaque detection technologies. Most of the preliminary development of algorithms for the detection and characterization of atherosclerotic plaques is performed using animal models of disease.6,24,25 It is possible that because of the biological differences that exist between animals and humans, the different tissue types contained in animal atherosclerotic lesions may not be similar to the lesions seen in human disease. However, distinct histological patterns such as the formation of fibrotic tissue or fibro-lipidic tissue do not differ much between species, and many other catheter-based technologies have been validated on the detection of these specific components in porcine models of disease.22,26
The inherent technological limitations of IVUS may have played an important role in our findings. It is possible that the early vascular lesions seen in our model could not be fully characterized because of the limited resolution of IVUS. In our experiments, although vascular lesions were identified in all the cases, it is possible that some of the plaque components may have been difficult to discern because of insufficient complexity of the lesions. However, the degree of precision needed to characterize plaque components linking these data to a clinical outcome remains unknown.
A major limitation needing to be addressed is the applicability of data that have been validated in ex vivo human tissue in detecting plaque components in an in vivo environment. Although standard definitions were used to analyze the data,11,13,20,21 the proper histological definitions that need to be included in the process of imaging validation have not been adequately defined. In addition, the potential artifacts in size and morphology created by handling and processing ex vivo samples needs to be taken into consideration at the time of tissue analysis. It is also theoretically possible that the acoustic signals acquired in an already processed tissue are different to the signals acquired in vivo, making the translation of data problematic.
In summary, to our knowledge, the present study represents the first in vivo correlation study of IVUS-VH with histology in an animal model of complex coronary lesions. In our study, IVUS-VH was not able to accurately detect the area of each specific component within the corresponding histological specimen. These findings highlight the importance of unifying criteria for histological definitions and animal model development. The development of an animal model with atherosclerotic lesions that share similar morphological features with human disease remains key for the further development of catheter based imaging modalities.
| Acknowledgments |
|---|
The research described in the manuscript was partially supported by a research grant from the Methodist DeBakey Heart Center and the Methodist Hospital Research Institute.
Disclosures
None.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. Alfonso and L. Hernando Intravascular ultrasound tissue characterization. I like the rainbow but... what's behind the colours? Eur. Heart J., July 2, 2008; 29(14): 1701 - 1703. [Full Text] [PDF] |
||||
![]() |
S. Voros Can Computed Tomography Angiography of the Coronary Arteries Characterize Atherosclerotic Plaque Composition?: Is the CAT (Scan) Out of the Bag? J. Am. Coll. Cardiol. Intv., April 1, 2008; 1(2): 183 - 185. [Full Text] [PDF] |
||||
![]() |
P. K. Shah Distal Embolization After Percutaneous Coronary Interventions: Prediction, Prevention, and Relevance J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1647 - 1648. [Full Text] [PDF] |
||||
![]() |
R. Kawaguchi, S. Oshima, M. Jingu, H. Tsurugaya, T. Toyama, H. Hoshizaki, and K. Taniguchi Usefulness of Virtual Histology Intravascular Ultrasound to Predict Distal Embolization for ST-Segment Elevation Myocardial Infarction J. Am. Coll. Cardiol., October 23, 2007; 50(17): 1641 - 1646. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Virmani and G. Nakazawa Animal Models and Virtual Histology Arterioscler. Thromb. Vasc. Biol., July 1, 2007; 27(7): 1666 - 1666. [Full Text] [PDF] |
||||
![]() |
S. K. Mehta, J. R. McCrary, A. D. Frutkin, W. J.S. Dolla, and S. P. Marso Intravascular ultrasound radiofrequency analysis of coronary atherosclerosis: an emerging technology for the assessment of vulnerable plaque Eur. Heart J., June 1, 2007; 28(11): 1283 - 1288. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
ATVB Home | Subscriptions | Archives | Feedback | Authors | Help | AHA Journals Home | Search Copyright © 2007 American Heart Association, Inc. All rights reserved. Unauthorized use prohibited. |